Tavsif

MV and HV insulation under full control

The S-25 DC device is universal high voltage tester. Capacitive loads, such as shielded power cables, will be properly and automatically discharged in case of an error or at the end of testing time by the integrated discharge device

Key switch, switch-on interlock and a protective ground circuit ensure maximum safety. 

S-25 DC stands out in particular. Not just with modest dimensions. Thanks to an internal rechargeable battery it enables mains-independent testing.

Sonel offers online training course on the use of this product (instructional video and consultation with a specialist).

Features

  • Easy operation
  • Robust construction, low weight
  • Integrated timer:
    • max. 30 min 
  • Discharge device:
    • integrated
  • Internal battery charger with deep discharge protection
  • Voltage measurement direct at HV output
  • Short-circuit-proof output:
    • electronically limited output current
  • Protective ground connection
  • Mains-independent by internal rechargeable battery

Fields of application

  • Testing at recommissioning of cable systems after maintenance and repair
  • Testing of newly installed cables an cable joints before commissioning
  • Voltage testing at cable sheath
  • Testing of electrical equipment

Xususiyatlar

Technical specification


S-25 DC Hipot S-50 DC Hipot S-80 DC Hipot S-110 DC Hipot S-120 DC Hipot
Power supply • internal rechargeable battery
• mains, switchable 115 / 230 V, 50-60 Hz
• external DC voltage source
mains 220…240 V, 50 / 60 Hz (optional: 115 V / 60 Hz mains 220…240 V, 50 / 60 Hz (optional: 115 V / 60 Hz mains 220…240 V, 50 / 60 Hz (optional: 115 V / 60 Hz mains 220…240 V, 50 / 60 Hz (optional: 115 V / 60 Hz
Power consumption max. 120 VA max. 900 VA max. 900 VA max. 900 VA max. 900 VA
Output voltage • 0…25 kV DC
• infinitely adjustable
• negative polarity
• ripple < 0,25%
• 0…50 kV DC
• infinitely adjustable
• negative polarity
• 0…80 kV DC
• infinitely adjustable
• negative polarity
• 0…110 kV DC
• infinitely adjustable
• negative polarity
• 0…120 kV DC
• infinitely adjustable
• negative polarity
Rated output current at max. output voltage 1.5 mA electronically limited 6 mA 5 mA 4 mA 3,5 mA
Battery operation at full load 45 min - - - -
Discharge - integrated automatic discharge device 3000 J 9.6 µF at 25 kV 7500 J 6 µF at 50 kV 12 250 J 4 µF at 75 kV 3.5 µF at 80 kV 15 000 J 3 µF at 100 kV 2.5 µF at 110 kV 18 750 J 2.4 µF at 125 kV 2.6 µF at 120 kV
Voltage measuring range 0…30 kV 0…50 kV 0…80 kV 0…110 kV 0…130 kV
Current measuring ranges 0…200 µA / 0…2 mA manual or automatic switching between ranges 0…100 μA / 1 mA / 10 mA manual switching between ranges 0…100 μA / 1 mA / 10 mA manual switching between ranges 0…100 μA / 1 mA / 10 mA manual switching between ranges 0…100 μA / 1 mA / 10 mA manual switching between ranges
Operating temperature -25…+55°C
-13…+131°F
-25…+55°C
-13…+131°F
-25…+55°C
-13…+131°F
-25…+55°C
-13…+131°F
-25…+55°C
-13…+131°F
Storage temperature -40…+70°C
-40…+158°F
-40…+70°C
-40…+158°F
-40…+70°C
-40…+158°F
-40…+70°C
-40…+158°F
-40…+70°C
-40…+158°F